A new calibration method for mixer delay measurements that requires no calibration mixer
A new calibration method for mixer delay measurements that requires no calibration mixer
复制标题
一种无需校准混频器的混频器延迟测量的新校准方法
DOI:
--
复制
发表时间:
2011
期刊:
影响因子:
--
通讯作者:
J. Dunsmore
中科院分区:
文献类型:
--
作者:
J. Dunsmore
Making phase and group delay response measurements on mixers and frequency converters has traditionally been a difficult and complicated process, requiring either modulated-signals (and demodulation), two-tone signals (a special form of modulation) or a system with a reference or reconverting mixer and a calibration or “golden” mixer. This paper presents a method to measure the phase response of a mixer on a VNA calibrating the input and output phase response independently, computing the transfer response directly using a phase reference, and is traceable to national standards. This method, for the first time, eliminates the need for any other mixer, reference or calibration, in the test system.