Cochlear implant electrode misplacement: incidence, evaluation, and management.
Cochlear implant electrode misplacement: incidence, evaluation, and management.
复制标题
人工耳蜗电极错位:发生率、评估和管理。
DOI:
10.1002/lary.23665
复制
发表时间:
2013
期刊:
影响因子:
--
通讯作者:
Williamson,RobertA
中科院分区:
文献类型:
--
作者:
Ying,Yu-LanMary;Lin,JerryW;Oghalai,JohnS;Williamson,RobertA
Objectives/HypothesisTo review the presentation and management of improper electrode array placement, and to help guide clinical decision‐making.Study DesignRetrospective case series.MethodsPediatric and adult cochlear implant patients managed from January 2001 to present whose electrode arrays were not placed properly within the cochlea or extended beyond the cochlea into the internal auditory canal or adjacent structures.ResultsFour patients, three pediatric and one adult, were identified from over 824 cases (< 1%) managed over the study duration. All cases had normal cochlear anatomy. These cases were initially identified due to poor auditory skill development or absent behavioral responses following implantation, which prompted imaging. Two patients presented several years after surgery. Sites of improper placement included the eustachian tube, vestibule, internal carotid artery canal, and internal auditory canal (IAC). Intraoperative findings and management are reviewed.ConclusionsElectrode array malpositioning is a rare, but serious and correctable complication in cochlear implant surgery. A multidisciplinary approach, including prompt audiologic evaluation and imaging, is important, particularly when benefit from the implant is limited or absent. Management of electrode arrays in the IAC may be more challenging. Laryngoscope, 2013