Intercomparison of a correlated-photon-based method to measure detector quantum efficiency.
Intercomparison of a correlated-photon-based method to measure detector quantum efficiency.
复制标题
用于测量探测器量子效率的基于相关光子的方法的相互比较。
DOI:
10.1364/ao.41.002914
复制
发表时间:
2002
期刊:
影响因子:
1.9
通讯作者:
M. Rastello
中科院分区:
文献类型:
--
作者:
A. Migdall;S. Castelletto;I. Degiovanni;M. Rastello
We report on the absolute calibration of photodetector quantum efficiency by using correlated photon sources, performed independently at two laboratories, the National Institute of Standards and Technology and the Istituto Elettrotecnico Nazionale (IEN). The goal is to use an interlaboratory comparison to demonstrate the inherent absoluteness of the photon correlation technique by showing its independence from the particular experimental setup. We find that detector nonuniformity limited this comparison rather than uncertainty inherent in the method itself. The ultimate goal of these investigations is development of a robust measurement protocol that allows the uncertainties of individual measurements to be determined experimentally and verified operationally. Furthermore, to demonstrate the generality of the procedure, the IEN measurement setup was also used to calibrate a fiber-coupled avalanche photodiode module. Uncertainties are evaluated for the detector both with and without fiber coupling and differences are discussed. The current IEN setup using a thinner and higher transmittance nonlinear crystal for the generation of correlated photons shows a significant improvement in overall accuracy with respect to previously reported results from IEN [Metrologia 32, 501-503 (1996)].