Noninvasive differentiation of uric acid versus non-uric acid kidney stones using dual-energy CT

Noninvasive differentiation of uric acid versus non-uric acid kidney stones using dual-energy CT
复制标题

DOI:
10.1016/j.acra.2007.09.016
复制
发表时间:
2007-12-01
期刊:
影响因子:
4.8
通讯作者:
McCollough, Cynthia H.
McCollough, Cynthia H.
中科院分区:
医学3区
文献类型:
--
作者:
Primak, Andrew N.;Fletcher, Joel G.;McCollough, Cynthia H.

文献摘要

被引文献

相似文献

基本原理和目标。确定商业实施产品中双能计算机断层扫描 (DECT) 区分尿酸 (UA) 结石与其他(非 UA)肾结石的准确性和灵敏度。材料和方法。将 40 颗含有尿酸 (n = 16)、羟基磷灰石 (n = 8)、草酸钙 (n = 8) 和胱氨酸 (n = 8) 的人类肾结石插入四个猪肾(每个 10 个)中,并放置在尸体脊柱前的 32 厘米水箱内。使用临床协议和自动曝光控制在双源 64 层计算机断层扫描 (CT) 系统上获得螺旋双能量扫描。在两种不同的准直(0.6 毫米和 1.2 毫米)和三种体模尺寸(中号、大号和超大号)内进行扫描,总共产生了六个图像数据集。使用 CT 系统上提供的双能软件工具对这些数据集进行分析,以确保准确性(正确分类为 UA 或非 UA 的结石数量)和灵敏度(对于 UA 结石)。结石特征与显微 CT 结果相关。对于中型和大型体模尺寸,DECT 技术显示出 100% 的准确度 (40/40),无论准直如何。对于超大模型尺寸和 0.6 毫米准直(导致噪声最大的数据集),无法对三块(两块胱氨酸和一块小 UA)结石进行分类(准确度为 93%,灵敏度为 94%)。对于超大体模尺寸和1.2毫米准直,双能工具未能识别两个小UA结石(准确度95%,灵敏度88%)。结论。在拟人体模模型中,双能CT可以准确区分尿酸结石和其他结石类型。
Rationale and Objectives. To determine the accuracy and sensitivity for dual-energy computed tomography (DECT) discrimination of uric acid (UA) stones from other (non-UA) renal stones in a commercially implemented product.Materials and Methods. Forty human renal stones comprising uric acid (n = 16), hydroxyapatite (n = 8), calcium oxalate (n = 8), and cystine (n = 8) were inserted in four porcine kidneys (10 each) and placed inside a 32-cm water tank anterior to a cadaver spine. Spiral dual-energy scans were obtained on a dual-source, 64-slice computed tomography (CT) system using a clinical protocol and automatic exposure control. Scanning was performed at two different collimations (0.6 mm and 1.2 mm) and within three phantom sizes (medium, large, and extra large) resulting in a total of six image datasets. These datasets were analyzed using the dual-energy software tool available on the CT system for both accuracy (number of stones correctly classified as either UA or non-UA) and sensitivity (for UA stones). Stone characterization was correlated with micro-CT.Results. For the medium and large phantom sizes, the DECT technique demonstrated 100% accuracy (40/40), regardless of collimation. For the extra large phantom size and the 0.6-mm collimation (resulting in the noisiest dataset), three (two cystine and one small UA) stones could not be classified (93% accuracy and 94% sensitivity). For the extra large phantom size and the 1.2-mm collimation, the dual-energy tool failed to identify two small UA stones (95% accuracy and 88% sensitivity).Conclusions. In an anthropomorphic phantom model, dual-energy CT can accurately discriminate uric acid stones from other stone types.