A NEW MULTIFREQUENCY MICROWAVE RADIOMETER FOR MEDICAL OPERATION
A NEW MULTIFREQUENCY MICROWAVE RADIOMETER FOR MEDICAL OPERATION
复制标题
DOI:
10.1016/0924-4247(92)80001-j
复制
发表时间:
1992-04-01
影响因子:
4.6
通讯作者:
BARDATI, F
中科院分区:
文献类型:
--
作者:
TOGNOLATTI, P;GIUSTO, R;BARDATI, F
Multi-frequency microwave radiometry has been considered for the measurement of sub-surface temperature distributions of human tissues, envisaged applications being mainly related to the non-invasive temperature monitoring in hyperthermia treatment of cancer. The sensor consists of a broadband microwave receiver, which generates a signal proportional to input radiation power, then a calibration processor averages the receiver output and computes an estimate of the input power in each frequency band. We have developed a new radiometer system of the radiation balance type to counteract the reflectivity errors which may affect the multi-frequency radiometric data to an intolerable amount. In this paper we discuss the reflectivity error of our system on the basis of a model of both incoming radiation and receiver. A formula, which propagates this error to temperature retrieval uncertainty, is also given.