Fast 3-D dark-field reflection-mode photoacoustic microscopy in vivo with a 30-MHz ultrasound linear array
Fast 3-D dark-field reflection-mode photoacoustic microscopy in vivo with a 30-MHz ultrasound linear array
复制标题
DOI:
10.1117/1.2976141
复制
发表时间:
2008-09-01
影响因子:
3.5
通讯作者:
Wang, Lihong V.
中科院分区:
文献类型:
--
作者:
Song, Liang;Maslov, Konstantin;Wang, Lihong V.
We present an in vivo dark-field reflection-mode photoacoustic microscopy system that performs cross-sectional (B-scan) imaging at 50 Hz with real-time beamforming and 3-D imaging consisting of 166 B-scan frames at 1 Hz with postbeamforming. To our knowledge, this speed is currently the fastest in photoacoustic imaging. A custom-designed light delivery system is integrated with a 30-MHz ultrasound linear array to realize dark-field reflection-mode imaging. Linear mechanical scanning of the array produces 3-D images. The system has axial, lateral, and elevational resolutions of 25, 70, and 200 mu m, respectively, and can image 3 mm deep in scattering biological tissues. Volumetric images of subcutaneous vasculature in rats are demonstrated in vivo. Fast 3-D photoacoustic microscopy is anticipated to facilitate applications of photoacoustic imaging in biomedical studies that involve dynamics and clinical procedures that demand immediate diagnosis. (C) 2008 Society of Photo-Optical Instrumentation Engineers. [DOI: 10.1117/1.2976141]