Spectral radiance calibrations between 165-300 nm: an interlaboratory comparison.
Spectral radiance calibrations between 165-300 nm: an interlaboratory comparison.
复制标题
165-300 nm 之间的光谱辐射率校准:实验室间比较。
DOI:
10.1364/ao.16.001788
复制
发表时间:
1977
期刊:
影响因子:
1.9
通讯作者:
D. Stuck
中科院分区:
文献类型:
--
作者:
J. Bridges;W. Ott;E. Pitz;A. Schulz;D. Einfeld;D. Stuck
The spectral radiance of deuterium lamps calibrated by the Max-Planck-Institut fuer Astronomie (MPI), by the U.S. National Bureau of Standards (NBS), and by the Physikalisch-Technische Bundesanstalt (PTB) are compared to check the agreement of UV radiometric scales. The NBS group used the optically thin continuum radiation from a wall-stabilized hydrogen arc as its fundamental radiometric standard, while the MPI and PTB groups used the synchrotron radiation facility in DESY. It is found that the spectral radiance scales based upon the DESY synchrotron and the NBS hydrogen arc are consistent, at least for one wavelength relative to another.