A probe-camera system for 3D ultrasound image reconstruction
A probe-camera system for 3D ultrasound image reconstruction
复制标题
用于 3D 超声图像重建的探头相机系统
DOI:
10.1007/978-3-319-67552-7_16
复制
发表时间:
2017
期刊:
影响因子:
--
通讯作者:
Koichi Ito
中科院分区:
文献类型:
--
作者:
Shuya Ito;Koichi Ito
This paper proposes a probe-camera system for 3D ultrasound (US) image reconstruction with probe-camera calibration and probe localization methods. The probe-camera calibration method employs an existing US phantom for convenience with a simple procedure. The probe localization method employs structure from motion (SfM) to estimate the camera motion. SfM is used to reconstruct 3D point clouds from multiple-view images and simultaneously estimate each camera position. Through experiments using the developed system, we demonstrate that the proposed method exhibits good performance to reconstruct 3D US volume.