Comparison of Spectra Optia and COBE Spectra apheresis systems' performances for red blood cell exchange procedures
Comparison of Spectra Optia and COBE Spectra apheresis systems' performances for red blood cell exchange procedures
复制标题
DOI:
10.1016/j.transci.2016.10.004
复制
发表时间:
2016-12-01
影响因子:
1.9
通讯作者:
Sarode, Ravi
中科院分区:
文献类型:
--
作者:
Kim, Jaehyup;Joseph, Ranjit;Sarode, Ravi
Background: Spectra Optia (Terumo BCT, Lakewood, CO) was FDA approved for red blood cell exchange (RBCx) procedures in January 2014 and is expected to replace COBE spectra (Terumo BCT) very soon in the USA. The performance characteristics of these devices for Isovolemic Hemodilution (IHD-RBCx) procedure were compared in this study.Methods: A total of 114 IHD-RBCx procedures from 19 patients were analyzed. For every patient, three procedures on each device with similar pre-procedure hematocrits were compared. Pre and post procedure laboratory parameters compared were hemoglobin S (HbS), hematocrits (Hct), platelet counts and fraction of cells remaining (FCR). Statistical analysis was performed using t-test adjusted by the Holm-Bonferroni method to reduce family-wise error rate.Results: There were no significant differences between these two devices in regards to HbS, Hct, FCR and platelet counts (p = > 0.05). However, rinseback volume (124.2 +/- 8.9 ml) and normal saline replacement volume during IHD phase (296.1 +/- 97.2 ml) were lower in Spectra Optia as compared to COBE Spectra (337 +/- 33.8 ml and 326.6 +/- 105.2 ml, p value