Evaluation of double photon coincidence Compton imaging method with GEANT4 simulation
Evaluation of double photon coincidence Compton imaging method with GEANT4 simulation
复制标题
利用GEANT4模拟评估双光子符合康普顿成像方法
DOI:
10.1016/j.nima.2017.05.039
复制
发表时间:
2017
期刊:
影响因子:
--
通讯作者:
H.
中科院分区:
文献类型:
--
作者:
Yoshihara;Y.;Shimazoe;K.;Mizumachi;Y.;& Takahashi;H.
Compton imaging has been used for various applications including astronomical observations, radioactive waste management, and biomedical imaging. The positions of radioisotopes are determined in the intersections of multiple cone traces through a large number of events, which reduces signal to noise ratio (SNR) of the images. We have developed an advanced Compton imaging method to localize radioisotopes with high SNR by using information of the interactions of Compton scattering caused by two gamma rays at the same time, as the double photon coincidence Compton imaging method. The targeted radioisotopes of this imaging method are specific nuclides that emit several gamma rays at the same time such as60Co,134Cs, and111In, etc. Since their locations are determined in the intersections of two Compton cones, the most of cone traces would disappear in the three-dimensional space, which enhances the SNR and angular resolution. In this paper, the comparison of the double photon coincidence Compton imaging method and the single photon Compton imaging method was conducted by using GEANT4 Monte Carlo simulation.