Image formation by nuclear magnetic resonance: The sensitive‐point method
Image formation by nuclear magnetic resonance: The sensitive‐point method
复制标题
核磁共振成像:敏感点法
DOI:
--
复制
发表时间:
1976
期刊:
影响因子:
--
通讯作者:
W. Hinshaw
中科院分区:
文献类型:
--
作者:
W. Hinshaw
An image formation method, using nuclear magnetic resonance techniques, has been developed which provides direct and detailed information on the interior details of heterogeneous samples. Time‐dependent magnetic field gradients are used to limit and control the region of spatial sensitivity of a standard pulsed NMR spectrometer. A continuous string of intense phase‐alternated rf pulses is used to provide a continuous monitor of the nuclear magnetization. The image is produced by directly plotting the phase‐detected resonance signal on a xy plotter as the limited region of spectrometer sensitivity is scanned through the sample. An analysis of the method in terms of the phenomenological Bloch equations, a description of the experimental method, and a representative selection of images are presented. A spatial resolution of about 0.3 mm has been achieved.