Evaluation of respiratory liver and kidney movements for MRI navigator gating.

Evaluation of respiratory liver and kidney movements for MRI navigator gating.
复制标题

DOI:
10.1002/jmri.22418
复制
发表时间:
2011-01
期刊:
Journal of magnetic resonance imaging : JMRI
影响因子:
--
通讯作者:
Hillenbrand CM
Hillenbrand CM
中科院分区:
其他
文献类型:
--
作者:
Song R;Tipirneni A;Johnson P;Loeffler RB;Hillenbrand CM

文献摘要

被引文献

相似文献

To determine the tracking factor by studying the relationship between kidney and diaphragm motions and to compare the efficiency of the gating-and-following and gating-only algorithms in reducing motion artifacts in navigator-gated scans. Diaphragm and kidney motions were measured by using real-time TrueFISP sequences from 10 healthy human volunteers in order to determine tracking factors at different acceptance windows. Mean tracking factors were used to calculate mean residual errors and improvement factors for the gating-and-following and gating-only algorithms. Mean tracking factors for ±4, ±6, ±8 mm and full acceptance windows ranged from 0.6 to 0.7, with large interindividual variations. Acceptance rates increased as the size of the acceptance window increased (acceptance rate for a 4 mm window ~50%). There was a greater reduction of motion errors by gating-and-following (maximum of 1.86 mm) than gating-only (maximum of 7.05 mm). Mean tracking factors obtained in this study can be used as a guideline for using the gating-and-following algorithm in navigator-gated kidney scans. The gating-and-following and gating-only algorithms were quantitatively compared, and it was found that the former is more effective in reducing motion errors.