Positional repeatability and variation in internal and external markers during volumetric-modulated arc therapy under end-exhalation breath-hold conditions for pancreatic cancer patients
Positional repeatability and variation in internal and external markers during volumetric-modulated arc therapy under end-exhalation breath-hold conditions for pancreatic cancer patients
复制标题
DOI:
10.1093/jrr/rraa054
复制
发表时间:
2020-09-01
影响因子:
2
通讯作者:
Sugimoto, Naozo
中科院分区:
文献类型:
--
作者:
Sasaki, Makoto;Nakamura, Mitsuhiro;Sugimoto, Naozo
The purpose of this study was to assess the positional repeatability of internal and external markers among multiple breath-hold (BH) sessions and evaluate the positional variation of these markers within BH sessions for volumetric-modulated arc therapy (VMAT) for pancreatic cancer patients. A total of 13 consecutive pancreatic cancer patients with an internal marker were enrolled. Single full-arc coplanar VMAT was delivered under end-exhalation BH conditions while monitoring the internal marker with kilovoltage (kV) X-ray fluoroscopy. Positional repeatability of the internal and external markers was determined by the difference between the reference and zero position in all BH sessions, and positional variation was defined by the displacement from the reference position in each BH session during megavolt beam delivery. The overall positional repeatability was 0.6 1.5 mm in the X-axis for the centroid of the internal marker (CoIM), -0.1 +/- 2.2 mm in the Y-axis for the CoIM, and 0.8 +/- 2.2 mm for the external marker. The frequency of an internal marker position appearing > 2 mm from the reference position in the Y-axis, despite the external marker position being