Development of Scanning-Type Synchrotron Radiation Mössbauer Microscope Using Focused X-Ray
Development of Scanning-Type Synchrotron Radiation Mössbauer Microscope Using Focused X-Ray
复制标题
开发使用聚焦X射线的扫描型同步辐射穆斯堡尔显微镜
DOI:
10.1143/jjap.43.389
复制
发表时间:
2004
影响因子:
1.5
通讯作者:
M. Seto
中科院分区:
文献类型:
--
作者:
T. Mitsui;Yasuhiro Kobayashi;M. Seto
A scanning-type synchrotron radiation Mössbauer microscope was designed and installed at the nuclear resonance beamline BL11XU of the SPring-8 facility. The focused Mössbauer probe beam (<φ20 µm) is obtained by the combination of a high-resolution monochromator, a multilayer X-ray focusing mirror and a pinhole slit. Using this Mössbauer microscope, we have performed, for the first time, the synchrotron radiation Mössbauer microscopic measurements of nuclear resonant forward scattering from an iron borate single crystal as a function of the sample position. It was found that the observed Mössbauer time spectra depend on the direction of the local hyperfine magnetic field and the local specimen thickness.