Personal cytometers: slow flow or no flow?
Personal cytometers: slow flow or no flow?
复制标题
个人细胞仪:慢流还是无流?
DOI:
10.1002/cyto.a.20284
复制
发表时间:
2006
期刊:
影响因子:
--
通讯作者:
Perlmutter,NancyG
中科院分区:
文献类型:
--
作者:
Shapiro,HowardM;Perlmutter,NancyG
Background: Although some manufacturers have optimistically described instruments with prices in the US 40,000rangeas“personalcytometers”,analogywiththepersonalcomputersuggeststhatthetargetpriceforatrue“personal”cytometershouldbeunder 5,000. Since such an apparatus could find a wide range of applications in cytomics in both developing and developed countries, it seemed desirable to consider its technical and economic feasibility. Methods: Using resolution targets and a variety of fluorescent bead standards immobilized on filters and/or slides, we evaluated high‐intensity LEDs as fluorescence excitation sources, relatively inexpensive CCD cameras as detectors, and 35 mm camera lenses and plastic low‐power microscope optics for light collection in a simple, inexpensive low‐resolution imaging cytometer. Results: The components tested could be combined toproduce an instrument capable of detecting fewer than 10,000 molecules of cell‐associated fluorescent label, and thus applicable to a broad range of cytometric tasks. Conclusions: Given the requirements for light sources, detectors, optics, mechanics, electronics and data analysis hardware and software, and the components presently available, it should be easier to reach the desired 5,000pricepointwithanimagecytometerthanwithaflowcytometer.©2006InternationalSocietyforAnalyticalCytology