Improved precision of absolute thermal-expansion measurements for ULE glass
Improved precision of absolute thermal-expansion measurements for ULE glass
复制标题
提高 ULE 玻璃绝对热膨胀测量的精度
DOI:
10.1117/12.258287
复制
发表时间:
1996
期刊:
影响因子:
--
通讯作者:
Deborah E. Morton
中科院分区:
文献类型:
--
作者:
M. J. Edwards;Emily H. Bullock;Deborah E. Morton
Improvements made in the ultrasonic thermal expansion measurement methods used at Corning Incorporated have significantly increased the precision of absolute CTE measurements on ULETM glass. Precise CTE data were required for modeling the performance of 8-meter mirror blanks. Several repeatability and reproducibility studies for many of the measurements performed throughout the 8- meter process were performed. Measurement equipment and procedures were then improved to obtain the required precision. A volume of absolute CTE data collected during the manufacture of three 8-meter mirror blanks is summarized, showing excellent results. Comparison with previously demonstrated precision is also given.