Intra and Inter-Observer Reliability of Mobile Tablet PACS Viewer System vs. Standard PACS Viewing Station-Diagnosis of Acute Central Nervous System Events

Intra and Inter-Observer Reliability of Mobile Tablet PACS Viewer System vs. Standard PACS Viewing Station-Diagnosis of Acute Central Nervous System Events
复制标题

移动平板电脑 PACS 查看器系统与标准 PACS 查看站的观察者内和观察者间可靠性 - 急性中枢神经系统事件的诊断

DOI:
--
复制
发表时间:
2013
期刊:
影响因子:
--
通讯作者:
Christopher Hancock
Christopher Hancock
中科院分区:
--
文献类型:
--
作者:
A. Bhatia;S. Patel;Gustavo Pantol;Yen;M. Plitnikas;Christopher Hancock

文献摘要

参考文献

被引文献

相似文献

背景和目的:苹果iPad的显示分辨率?为1024 × 768像素,大于 生成典型的CT或MRI图像所需的时间。本研究的目的 是确定特定的CT和MR序列是否可以准确地解释 在移动终端/PACS软件平台上, 固定高分辨率监视器/PACS放射工作站。如果是这样, 允许放射科医生提供类似的解释, 在成像中心或医院。材料 和方法:本研究是由研究者发起的,单中心, 回顾性、非随机、IRB批准的研究。包括5名放射科医生 本研究每个人都独立解释了特定的CT和MR序列, 传统的高分辨率液晶显示器通过eFilm?软件以及 iPad?使用Osirix?的移动终端软件程序。重复解释是 进行,每次判读之间至少间隔4周 病人本研究包括:50例CTA灌注成像患者, 50名患者接受了脑部MRI检查,50名患者接受了脊柱MRI检查, 是通过急诊室请求生成的图像研究订单。 随后,每个放射科医生对每个患者的解释结果是 统计学比较,以评价观察者内和观察者间 可靠性结果:CTA灌注脑研究内的参数设置 表现出良好的观察者内变异性。所有的参数都在 MRI脑部研究表明,观察者内变异性极好, 科恩kappa值> 0.75。科恩的卡帕值为董事会认证 神经放射科医生证明了所有参数的出色变异性; 住院放射科医师具有良好的变异性, 0.75附近结论:数据和统计分析表明,便携式移动的 比如苹果iPad?能显示足够的CT和MRI分辨率 序列,以准确诊断急性中枢神经系统损伤, 其他非急性病理。
Background and Purpose: The display resolution of the Apple iPad? is 1024 × 768 pixels, which is greater than that required for generating the typical CT or MRI images. The purpose of this study is to determine if specific CT and MR sequences can be interpreted accurately on mobile device/PACS software platforms when compared to a traditional stationary high resolution monitor/PACS radiological workstation. If so, this allows radiologists to provide comparable interpretation as if they were onsite at an imaging center or hospital. Materials and Methods: This study is an investigator initiated, single site, retrospective, nonrandomized, IRB approved study. Five radiologists were included in this study. Each independently interpreted specific CT and MR sequences on traditional high-resolution LCD monitors via eFilm? software as well as an iPad? mobile device using Osirix? software program. Repeat interpretations were performed, with 4 weeks minimum interval between interpretations of each patient. This investigation included: 50 patients with CTA perfusion imaging, 50 patients with MRI of the brain, and 50 patients with MRI of the spine, which were image study orders generated through emergency room requests. Subsequently, interpretive results of each radiologist for each patient were statistically compared to evaluate for intra-observer and inter-observer reliability. Results: The parameters set within the CTA perfusion brain studies demonstrated excellent intra-observer variability. All of the parameters within the MRI brain studies demonstrated excellent intra-observer variability with a Cohen’s kappa value > 0.75. The Cohen’s kappa values for the board certified neuroradiologist demonstrated excellent variability for all parameters; the resident radiologists had good variability, with a majority of kappa values near 0.75. Conclusions: The data and statistical analysis demonstrated that portable mobile devices such as the Apple iPad? can display adequate resolution of CT and MRI sequences to accurately diagnose acute central nervous system injuries and other non-acute pathology.
显示图像阈值和观察者检测性能。