Microdialysis sampling and chemiluminescence detection for in vivo and real-time study of the lead metabolism in rabbit blood
Microdialysis sampling and chemiluminescence detection for in vivo and real-time study of the lead metabolism in rabbit blood
复制标题
DOI:
10.1007/s00604-006-0699-2
复制
发表时间:
2007-07
影响因子:
5.7
通讯作者:
X. Hou;Z. Zhang;Yan Zhao;Jie Ma
中科院分区:
文献类型:
--
作者:
X. Hou;Z. Zhang;Yan Zhao;Jie Ma
The chemiluminescence (CL) system K2MnO4-luminol is shown to be useful for the determination of lead(II). The method is based on the catalytic effect of Pb(II) on the CL reaction. The linear range was 3 × 10−3–9 × 10−1mg L−1(r= 0.9971) and the relative standard deviation (R.S.D.) for 5 × 10−2mg L−1Pb(II) measurement was 0.7% (n= 11). The detection limit was 3 × 10−4mg L−1(3σ) Pb(II). Based on this, an in vivo, on-line, real-time analytical system for monitoring the metabolism of free lead(II) in rabbit blood was developed. A microdialysis probe, implanted in the vein of a rabbit, was perfused with perfusate at a flow rate of 5 µL min−1. The concentration of free lead(II) in the dialysate was determined on-line with a flow-injection CL system. This system included microdialysis sampling, on-line separation and chemiluminescence detection. The concentration-time curve of lead(II) was in accordance with the one-compartmental open model, T1/2(elimination half-life), Tmax(peak time) and Cmax(peak concentration) were 37.77 min, 85.20 min and 0.137 mg L−1, respectively.