Comparison of radiation doses from multislice computed tomography coronary angiography and conventional diagnostic angiography
Comparison of radiation doses from multislice computed tomography coronary angiography and conventional diagnostic angiography
复制标题
DOI:
10.1016/j.jacc.2005.11.078
复制
发表时间:
2006-05-02
影响因子:
24
通讯作者:
Baumbach, A
中科院分区:
文献类型:
--
作者:
Coles, DR;Smail, MA;Baumbach, A
OBJECTIVES The aim of this study was to quantify and compare effective doses from conventional angiography and multislice computed tomography (MSCT) coronary angiography using a 16-slice scanner.BACKGROUND Multislice computed tomography is now a viable modality for cardiac imaging. However, for any diagnostic use of ionizing radiation, the risk to the patient must be considered and justified.METHODS Multislice computed tomography angiography and conventional angiography were used to assess 180 patients with suspected coronary artery disease. Estimates of effective dose were derived from exposure data recorded for each patient examination. For each modality, a comparable calculation technique was used, based on Monte Carlo modeling of the standard Cristy phantom.RESULTS In a subset of 91 directly comparable patients the mean effective dose for MSCT coronary angiography was 14.7 mSv (SD 2.2) and that for conventional angiography was 5.6 mSv (SD 3.6). A significant difference in effective dose was seen between the two protocols.CONCLUSIONS The mean effective dose for MSCT coronary angiography was significantly higher than that for conventional angiography. As MSCT cardiac scanners become increasingly available, operators must be aware of the radiation dose and the factors that affect it.