Determination of Lead with a Carbon Paste Electrode Modified with a Chelating Resin in Rainwater and Human Urine
Determination of Lead with a Carbon Paste Electrode Modified with a Chelating Resin in Rainwater and Human Urine
复制标题
用螯合树脂修饰的碳糊电极测定雨水和人尿中的铅
DOI:
10.1080/03067318908026890
复制
发表时间:
1989
影响因子:
2.6
通讯作者:
L. Hernandez
中科院分区:
文献类型:
--
作者:
M. S. Escribano;J. R. Procopio;J. P. Macias;L. Hernandez
Abstract A method is described for determination of lead using a carbon paste electrode modified wtih Dowex A-1 chelating ion-exchanger resin and cyclic voltammetry for the electrochemical measurement. The best conditions for preconcentration were found to be open circuit for 10 min in ammonium acetate, pH 5. Optima parameters for the stripping measurements were a HCl 0.1 M solution as supporting electrolyte, a delay time of 48s at –1.0 vs. S.C.E. and a scan rate of 150mv/s. The anodic peak at –0.52V vs. S.C.E. permits adequate quantification of lead in concentration between 4 and 200 ng/ml with a LD (3[sgrave]) of 1.2ng/ml. The method was used for determination of lead in rainwater and human urine.