Closed-Loop Control in Fused MR-TRUS Image-Guided Prostate Biopsy
Closed-Loop Control in Fused MR-TRUS Image-Guided Prostate Biopsy
复制标题
融合 MR-TRUS 图像引导前列腺活检中的闭环控制
DOI:
--
复制
发表时间:
2007
期刊:
影响因子:
--
通讯作者:
B. Wood
中科院分区:
文献类型:
--
作者:
Sheng Xu;J. Kruecker;P. Guion;N. Glossop;Z. Neeman;P. Choyke;Anurag K. Singh;B. Wood
Multi-modality fusion imaging for targeted prostate biopsy is difficult because of prostate motion during the biopsy procedure. A closed-loop control mechanism is proposed to improve the efficacy and safety of the biopsy procedure, which uses real-time ultrasound and spatial tracking as feedback to adjust the registration between a preoperative 3D image (e.g. MRI) and real-time ultrasound images. The spatial tracking data is used to initialize the image-based registration between intraoperative ultrasound images and a preoperative ultrasound volume. The preoperative ultrasound volume is obtained using a 2D sweep and manually registered to the MRI dataset before the biopsy procedure. The accuracy of the system is 2.3 +/- 0.9 mm in phantom studies. The results of twelve patient studies show that prostate motion can be effectively compensated using closed-loop control.