Physical activity energy expenditure measured using a portable telemetric device in comparison with a mass spectrometer
Physical activity energy expenditure measured using a portable telemetric device in comparison with a mass spectrometer
复制标题
使用便携式遥测设备与质谱仪测量的身体活动能量消耗
DOI:
--
复制
发表时间:
2003
影响因子:
18.4
通讯作者:
A. Lorenzo
中科院分区:
文献类型:
--
作者:
C. Maiolo;G. Melchiorri;L. Iacopino;S. Masala;A. Lorenzo
Background: Several studies have evaluated the accuracy of portable metabolic gas analysis systems, which measure physical activity energy expenditure. Since the Cosmed K2 telemetric system, two K4 devices (RQ and b2) have been developed. Objective: To compare measurements using the Cosmed K4 RQ with measurements by a traditional method (mass spectrometer) for a sample of healthy subjects performing physical exercise of various intensities. Methods: Nine healthy male footballers (mean (SD) age 18.3 (2.2) years) performed an incremental exercise running test from 8 km/h to exhaustion, with an increase of 2 km/h for each level of the test protocol. The expired air was collected simultaneously using the Cosmed K4 RQ and Airspec QP9000 mass spectrometer. Oxygen consumption (Vo2) and carbon dioxide production (Vco2) were measured at rest and at each exercise intensity. Results: No significant differences were found between Vo2 and Vco2 measured by the two instruments, at rest or during exercise. Conclusions: The application of the Cosmed K4 RQ system for Vo2 and Vco2 measurements at rest and during both submaximal and maximal exercise may be of interest to sport physiologists, nutritionists, and doctors.