Comparison of SPAMM and SENC Methods for Evaluating Peak Circumferential Strain at 3T

Comparison of SPAMM and SENC Methods for Evaluating Peak Circumferential Strain at 3T
复制标题

DOI:
10.2463/mrms.2012-0045
复制
发表时间:
2013-01-01
影响因子:
3
通讯作者:
Sakata, Motomichi
Sakata, Motomichi
中科院分区:
医学4区
文献类型:
--
作者:
Sugimori, Hiroyuki;Oyama-Manabe, Noriko;Sakata, Motomichi

文献摘要

被引文献

相似文献

我们比较了峰值周向应变(Ecc)值与空间磁化调制(SPAMM)和应变编码(SENC)磁共振(MR)成像在3特斯拉。两种方法的平均Ecc峰值的相关系数有统计学意义。然而,SPAMM的平均峰值Ecc值(-13.5% +/- 3.3%)显著低于SENC(-21.6% +/- 3.4%)(P < 0.0001)。在3 T下,SENC方法显示出比SPAMM方法更高的周向应变。
We compared peak circumferential strain (Ecc) values with spatial modulation of magnetization (SPAMM) and strain-encoded (SENC) magnetic resonance (MR) imaging at 3 tesla. Correlation coefficients of the averaged peak Ecc values of the 2 methods were statistically significant. However, the average peak Ecc value was significantly lower with SPAMM (-13.5% +/- 3.3%) than with SENC (-21.6% +/- 3.4%) (P < 0.0001). The SENC method showed higher circumferential strain than the SPAMM method at 3T.