Calibration of a thermocouple for measurement of oscillating temperature
Calibration of a thermocouple for measurement of oscillating temperature
复制标题
DOI:
10.1063/1.2140489
复制
发表时间:
2005-12
影响因子:
1.6
通讯作者:
Y. Tashiro;T. Biwa;T. Yazaki
中科院分区:
文献类型:
--
作者:
Y. Tashiro;T. Biwa;T. Yazaki
We report on the dynamic calibration of a thermocouple for the measurement of the oscillating temperature. Temperature oscillation is induced in a gas-filled tube by a periodically forced pressure oscillation and measured by a thermocouple. The radial profile of the measured temperature oscillations is compared with the theoretical one, which is determined from the simultaneously measured pressure. A response function of the thermocouple is obtained from the difference in amplitude and phase angle between them by varying the diameter of the thermocouple, oscillating frequency, tube radius, and working gas. We can obtain a true temperature oscillation by using the response function given in this experiment.