Improved Detection of Sentinel Lymph Nodes in SPECT/CT Images Acquired Using a Low- to Medium-Energy General-Purpose Collimator

Improved Detection of Sentinel Lymph Nodes in SPECT/CT Images Acquired Using a Low- to Medium-Energy General-Purpose Collimator
复制标题

DOI:
10.1097/rlu.0b013e31828da362
复制
发表时间:
2014-01-01
影响因子:
10.6
通讯作者:
Kinuya, Seigo
Kinuya, Seigo
中科院分区:
医学3区
文献类型:
--
作者:
Yoneyama, Hiroto;Tsushima, Hiroyuki;Kinuya, Seigo

文献摘要

被引文献

相似文献

目的:使用低能量高分辨率(LEHR)准直器进行淋巴管造影会导致注射部位出现星形伪影。本研究的目的是确认是否低到中等能量的通用(LMEGP)准直器的分辨率较低的补偿间隔渗透和减少星形artifaction.Methods的程度减少:共106例女性乳腺癌患者,活检诊断,参加了这项研究。肿瘤周围注射Tc-99 mPhytate(37 MBq,1 mCi),3 ~ 4 h后行平面及SPECT/CT显像。结果:LEHR和LEHR + LMEGP准直器的平面显像分别在96.2%(102/106)和99.1%(105/106)的患者中检出前哨淋巴结(SLN)。使用LEHR和LEHR + LMEGP准直器的平面和SPECT/CT成像相结合,SLN的阳性识别率分别为97.2%(103/106)和100%(106/106)。LMEGP准直器的使用改善了SLN的检测。
Purpose: The use of the low-energy high-resolution (LEHR) collimator for lymphoscintigraphy causes the appearance of star-shaped artifacts at injection sites. The aim of this study was to confirm whether the lower resolution of the low-to medium-energy general-purpose (LMEGP) collimator is compensated by decrease in the degree of septal penetration and the reduction in star-shaped artifacts.Methods: A total of 106 female patients with breast cancer, diagnosed by biopsy, were enrolled in this study. Tc-99m phytate (37 MBq, 1 mCi) was injected around the tumor, and planar and SPECT/CT images were obtained after 3 to 4 hours. When sentinel lymph nodes (SLNs) could not be identified from planar and SPECT/CT images by using the LEHR collimator, we repeated the study with the LMEGP collimator.Results: Planar imaging performed using the LEHR and LEHR + LMEGP collimators positively identified SLNs in 96.2% (102/106) and 99.1% (105/106) of the patients, respectively. Using combination of planar and SPECT/CT imaging with the LEHR and LEHR + LMEGP collimators, SLNs were positively identified in 97.2% (103/106) and 100% (106/106) of the patients, respectively.Conclusions: The LMEGP collimator provided better results than the LEHR collimator because of the lower degree of septal penetration. The use of the LMEGP collimator improved SLN detection.