IMPROVED SENSITIVITY IN FLOW CYTOMETRIC INTRACELLULAR IONIZED CALCIUM MEASUREMENT USING FLUO-3 FURA RED FLUORESCENCE RATIOS
IMPROVED SENSITIVITY IN FLOW CYTOMETRIC INTRACELLULAR IONIZED CALCIUM MEASUREMENT USING FLUO-3 FURA RED FLUORESCENCE RATIOS
复制标题
DOI:
10.1002/cyto.990170205
复制
发表时间:
1994-10-01
期刊:
影响因子:
--
通讯作者:
RABINOVITCH, PS
中科院分区:
文献类型:
--
作者:
NOVAK, EJ;RABINOVITCH, PS
Measurement of changes in intracellular ionized calcium concentrations ([Ca2+](i)) has proved to be of wide use in the study of cellular responses to activating stimuli. The fluorescent dye Indo-1 has successfully been used in flow cytometry for this purpose, and when used as a ratiometric indicator it provides optimum sensitivity and accuracy. Unfortunately, this dye requires ultraviolet (UV) excitation which is often not available. We show here that similar results can be obtained using a ratio of green to red fluorescence from the simultaneous loading of the dyes Fura Red and fluo-3. Both Fura Red and fluo-3 are excited using the commonly available blue 488 nm laser Line. With appropriate concentrations of the two dyes, the magnitude of response with the fluo-3/Fura Red ratio is greater than that achieved with indo-1, while the intercellular variation in measurement is similar to that seen with indo-1. Analyses can be simultaneously combined with immunofluorescent detection of PE-labeled antibodies to enable [Ca2+](i) measurement within cell subsets. (C) 1994 Wiley-Liss, Inc.