Smartphone-based chemiluminescence detection of aflatoxin B1 via labelled and label-free dual sensing systems.
Smartphone-based chemiluminescence detection of aflatoxin B1 via labelled and label-free dual sensing systems.
复制标题
DOI:
10.1016/j.foodchem.2023.135654
复制
发表时间:
2023-02
期刊:
影响因子:
8.8
通讯作者:
Junning Ma;Y. Guan;F. Xing;Yan Wang;Xu Li;Qiang Yu;Xiao-Bo Yu
中科院分区:
文献类型:
--
作者:
Junning Ma;Y. Guan;F. Xing;Yan Wang;Xu Li;Qiang Yu;Xiao-Bo Yu
To develop a sensing platform for onsite determination of AFB1in foodstuffs, we developed smartphone-based chemiluminescence detection of AFB1via labelled and label-free dual modes. The labelled mode was characteristic of double streptavidin–biotin mediated signal amplification, obtaining limit of detection (LOD) of 0.04 ng/mL in the linear range of 1–100 ng/mL. To reduce the complexity in the labelled system, a label-free mode based on both split aptamer and split DNAzyme was fabricated. A satisfactory LOD of 0.33 ng/mL was generated in the linear range of 1–100 ng/mL. Both labelled and label-free sensing systems achieved outstanding recovery rate in AFB1-spiked maize and peanut kernel samples. Finally, two systems were successfully integrated into smartphone-based portable device based on custom-made components and android application, achieving comparable AFB1detection ability to a commercial microplate reader. Our systems hold huge potential for AFB1onsite detection in food supply chain.