TIME-RESOLVED FLOW-CYTOMETRY FOR THE MEASUREMENT OF LANTHANIDE CHELATE FLUORESCENCE .1. CONCEPT AND THEORETICAL EVALUATION
TIME-RESOLVED FLOW-CYTOMETRY FOR THE MEASUREMENT OF LANTHANIDE CHELATE FLUORESCENCE .1. CONCEPT AND THEORETICAL EVALUATION
复制标题
DOI:
10.1002/cyto.990160302
复制
发表时间:
1994-07-01
期刊:
影响因子:
--
通讯作者:
ANLIKER, M
中科院分区:
文献类型:
--
作者:
CONDRAU, MA;SCHWENDENER, RA;ANLIKER, M
The concept of a flow cytometer suited for the time-resolved measurement of lanthanide chelate luminescence with a decay time on the order of 10 mu s to 2 ms is presented and evaluated. The instrument proposed encompasses a continuous-wave laser for fluorescence excitation and an optical switch for the elimination of cellular autofluorescence decaying within 1 ns to 1 mu s during the luminescence detection period. The slowly decaying fluorescence is to be quantified by a photon-counting system, whereas light scatter and prompt fluorescence parameters are acquired by a conventional detection system.The detection limit of the method, in terms of the smallest detectable number of fluorescing chelates per cell, is examined. It was found to be nearly 30,000 complexes of a europium chelate with a decay time of 1.6 ms and a quantum efficiency of 17%, independent of fast decaying cellular autofluorescence or prompt dye emission intensity. The probability of cells passing through the instrument without being detected while the laser beam is turned off was estimated, and the implications for cell throughput and sorting performance of the instrument were assessed. At typical fluorescence detection intervals of 500 mu s to 1 ms, cell flow rates of 100-200 particles per second lead to detection probabilities of more than 90% and sorting purities comparable to those found in conventional fluorescence-activated cell sorting. (C) 1994 Wiley-Liss, Inc.