MR imaging of the fetal brain at 1.5T and 3.0T field strengths: comparing specific absorption rate (SAR) and image quality.
MR imaging of the fetal brain at 1.5T and 3.0T field strengths: comparing specific absorption rate (SAR) and image quality.
复制标题
DOI:
10.1515/jpm-2014-0268
复制
发表时间:
2015-03
影响因子:
2.4
通讯作者:
Romero R
中科院分区:
文献类型:
--
作者:
Krishnamurthy U;Neelavalli J;Mody S;Yeo L;Jella PK;Saleem S;Korzeniewski SJ;Cabrera MD;Ehterami S;Bahado-Singh RO;Katkuri Y;Haacke EM;Hernandez-Andrade E;Hassan SS;Romero R
1) To evaluate the feasibility of fetal brain magnetic resonance imaging (MRI) using a fast spin echo sequence at 3.0T field strength with low radio frequency (rf) energy deposition (as measured by specific absorption rate – SAR) than 1.5T; 2) and to compare image quality, tissue contrast and conspicuity between 1.5T and 3.0T MRI. T2 weighted images of the fetal brain at 1.5T were compared to similar data obtained in the same fetus using a modified sequence at 3.0T. Quantitative whole body SAR and normalized image signal to noise ratio (SNR); and a nominal scoring scheme based evaluation of diagnostic image quality and tissue contrast & conspicuity for specific anatomical structures in the brain, were compared between 1.5T and 3.0T. 12 pregnant women underwent both 1.5T and 3.0T MRI examinations. The image SNR was significantly higher (p=0.03) and whole body SAR was significantly lower (p < 0.0001) for images obtained at 3.0T compared to 1.5T. All cases at both field strengths were scored as having diagnostic image quality. Images from 3.0T MRI (compared to 1.5T) were: 1) equal (57%; 21/37) or superior (35%; 13/37) for tissue contrast; and 2) equal (61%; 20/33) or superior (33%, 11/33) for conspicuity. It is possible to obtain fetal brain images with higher resolution and better SNR at 3.0T with simultaneous reduction in SAR compared to 1.5T. Images of the fetal brain obtained at 3.0T demonstrated superior tissue contrast and conspicuity compared to 1.5T.