A 128-Channel High Precision Time Measurement Module
A 128-Channel High Precision Time Measurement Module
复制标题
DOI:
10.2478/mms-2013-0024
复制
发表时间:
2013-06
影响因子:
1
通讯作者:
L. Kang;Lei Zhao;Jiawen Zhou;Shubin Liu;Q. An
中科院分区:
文献类型:
--
作者:
L. Kang;Lei Zhao;Jiawen Zhou;Shubin Liu;Q. An
In the external target experiment for heavy ion collisions in the HIRFL -CSR, Multi-Wire Drift Chambers are used to measure the drift time of charged particles to obtain the track information. This 128 -channel high precision time measurement module is designed to perform the time digitization. The data transfer is based on a PXI interface to guarantee a high data rate. Test results show that a 100 ps resolution with a data transfer rate up to 40 MBps has been achieved; this module has also been proven to function well with the detector through a commissioning test. Keyword : high precision, time measurement, PXI, data transfer, commissioning test.