A microfluidic chip for heterogeneous immunoassay using electrokinetical control
A microfluidic chip for heterogeneous immunoassay using electrokinetical control
复制标题
DOI:
10.1007/s10404-005-0040-0
复制
发表时间:
2005-10-01
影响因子:
2.8
通讯作者:
Li, DQ
中科院分区:
文献类型:
--
作者:
Hu, GQ;Gao, YL;Li, DQ
This article presents the development of a novel, automated, electrokinetically controlled heterogeneous immunoassay on a poly(dimethylsiloxane) (PDMS) microfluidic chip. A numerical method has been developed to simulate the electrokinetically driven, time-dependent delivery processes of reagents and washing solutions within the complex microchannel network. Based on the parameters determined from the numerical simulations, fully automated on-chip experiments to detect Helicobacter pylori were accomplished by sequentially changing the applied electric fields. Shortened assay time and much less reagent consumptions are achieved by using this microchannel chip while the detection limit is comparable to the conventional assay. There is a good agreement between the experimental result and numerical prediction, demonstrating the effectiveness of using CFD to assist the experimental studies of microfluidic immunoassay.