Fixed gantry tomosynthesis system for radiation therapy image guidance based on a multiple source x-ray tube with carbon nanotube cathodes

Fixed gantry tomosynthesis system for radiation therapy image guidance based on a multiple source x-ray tube with carbon nanotube cathodes
复制标题

DOI:
10.1118/1.3110067
复制
发表时间:
2009-05-01
期刊:
影响因子:
3.8
通讯作者:
Bani-Hashemi, Ali R.
Bani-Hashemi, Ali R.
中科院分区:
医学3区
文献类型:
--
作者:
Maltz, Jonathan S.;Sprenger, Frank;Bani-Hashemi, Ali R.

文献摘要

被引文献

相似文献

The authors present the design and simulation of an imaging system that employs a compact multiple source x-ray tube to produce a tomosynthesis image from a set of projections obtained at a single tube position. The electron sources within the tube are realized using cold cathode carbon nanotube technology. The primary intended application is tomosynthesis-based 3D image guidance during external beam radiation therapy. The tube, which is attached to the gantry of a medical linear accelerator (linac) immediately below the multileaf collimator, operates within the voltage range of 80-160 kVp and contains a total of 52 sources that are arranged in a rectilinear array. This configuration allows for the acquisition of tomographic projections from multiple angles without any need to rotate the linac gantry. The x-ray images are captured by the same amorphous silicon flat panel detector employed for portal imaging on contemporary linacs. The field of view (FOV) of the system corresponds to that part of the volume that is sampled by rays from all sources. The present tube and detector configuration provides an 8x8 cm(2) FOV in the plane of the linac isocenter when the 40.96x40.96 cm(2) imaging detector is placed 40 cm from the isocenter. Since this tomosynthesis application utilizes the extremities of the detector to record image detail relating to structures near the isocenter, simultaneous treatment and imaging is possible for most clinical cases, where the treated target is a small region close to the linac isocenter. The tomosynthesis images are reconstructed using the simultaneous iterative reconstruction technique, which is accelerated using a graphic processing unit. The authors present details of the system design as well as simulated performance of the imaging system based on reprojections of patient CT images.