Response of the alanine/ESR dosimetry system to MV x-rays relative to 60Co radiation
Response of the alanine/ESR dosimetry system to MV x-rays relative to 60Co radiation
复制标题
丙氨酸/ESR 剂量测定系统对 MV X 射线相对于 60Co 辐射的响应
DOI:
--
复制
发表时间:
2008
影响因子:
3.5
通讯作者:
F. Renner
中科院分区:
文献类型:
--
作者:
M. Anton;R. Kapsch;M. Krystek;F. Renner
The measurand relevant in dosimetry for radiation therapy is the absorbed dose to water, DW. The Physikalisch-Technische Bundesanstalt (PTB) has established a secondary standard for DW for high-energy photon and electron radiation based on electron spin resonance (ESR) of the amino acid alanine. Since the calibration is usually performed using 60Co radiation while a huge part of the external radiation therapy is done with high-energy x-rays from linear accelerators, determination of the response is an important issue. The results presented in this paper are the most accurate ones available today with uncertainties assigned to the relative response for 8 MV and 16 MV of the order of 0.3%. The experimental results are compared to Monte Carlo simulations using the EGSnrc software package. In the appendix, it is demonstrated how mean values from repetitive irradiations and their uncertainties are obtained in a consistent way using Bayesian statistics, even in the presence of at first sight inconsistent data. It is important to note that the formulae derived to obtain the final results follow from first principles, without recurring to ad hoc solutions or simple recipes and are valid for all kinds of repetitive measurements.