Shortening MR image acquisition time for volumetric interpolated breath-hold examination with a recently developed parallel imaging reconstruction technique: Clinical feasibility

Shortening MR image acquisition time for volumetric interpolated breath-hold examination with a recently developed parallel imaging reconstruction technique: Clinical feasibility
复制标题

DOI:
10.1148/radiol.2302021230
复制
发表时间:
2004-02-01
期刊:
影响因子:
19.7
通讯作者:
Rofsky, NM
Rofsky, NM
中科院分区:
医学1区
文献类型:
--
作者:
McKenzie, CA;Lim, D;Rofsky, NM

文献摘要

被引文献

相似文献

最近开发的并行磁共振 (MR) 成像技术(在 k 空间中具有增强半径的并行成像)被用来加速对 20 名转诊进行临床肝脏成像的患者进行的容积插值屏气检查 (VIBE)。还采集了这些患者的非加速 MR 图像。使用五分制对图像质量进行评分。加速导致图像质量下降:非加速图像的图像质量明显更高(P
A recently developed parallel magnetic resonance (MR) imaging technique, parallel imaging with an augmented radius in k space, was used to accelerate the volumetric interpolated breath-hold examination (VIBE) performed in 20 patients referred for clinical liver imaging. Nonaccelerated MR images were also acquired in these patients. A five-point scale was used to score the quality of the images. The acceleration resulted in reduced image quality: The nonaccelerated images had a significantly higher (P