Biosensor arrays for simultaneous measurement of glucose, lactate, glutamate, and glutamine
Biosensor arrays for simultaneous measurement of glucose, lactate, glutamate, and glutamine
复制标题
DOI:
10.1016/s0956-5663(01)00298-6
复制
发表时间:
2002-04-01
影响因子:
12.6
通讯作者:
Urban, GA
中科院分区:
文献类型:
--
作者:
Moser, I;Jobst, G;Urban, GA
For simultaneous measurement of glucose, lactate, glutamine, and glutamate a biosensor array is implemented in a micro flow-system thus giving a microsystem. The microsystem consists of a glass chip with the integrated biosensor array and a bottom part, which comprises a gold counter electrode, a 300 m thick seal, and electrical interconnection lines. The flow device has a total internal volume of 2.1 or 6 mul when integrated with a mixer on chip. The biosensors with no crosstalking and high long term stability were produced by modifying the electrochemical transducers and utilizing photopatternable enzyme membranes. The use of appropriate miniaturization technology leads to mass producable devices for in vivo and ex vivo applications in whole blood and fermentation broth. Due to a novel glutaminase with an activity optimum in the neutral pH range direct and simultaneous monitoring of glutamine together with glucose, lactate, and glutamate could be performed. (C) 2002 Elsevier Science B.V. All rights reserved.