Platelet adhesion to polyurethane urea under pulsatile flow conditions.

Platelet adhesion to polyurethane urea under pulsatile flow conditions.
复制标题

脉动流条件下血小板对聚氨酯脲的粘附。

DOI:
10.1111/aor.12296
复制
发表时间:
2014
期刊:
影响因子:
2.4
通讯作者:
Manning,KeefeB
Manning,KeefeB
中科院分区:
工程技术3区
文献类型:
--
作者:
Navitsky,MichaelA;Taylor,JoshuaO;Smith,AlexanderB;Slattery,MargaretJ;Deutsch,Steven;Siedlecki,ChristopherA;Manning,KeefeB

文献摘要

被引文献

相似文献

Platelet adhesion to a polyurethane urea surface is a precursor to thrombus formation within blood‐contacting cardiovascular devices, and platelets have been found to adhere strongly to polyurethane surfaces below a shear rate of approximately 500 s−1. The aim of the current work is to determine the properties of platelet adhesion to the polyurethane urea surface as a function of time‐varying shear exposure. A rotating disk system was used to study the influence of steady and pulsatile flow conditions (e.g., cardiac inflow and sawtooth waveforms) for platelet adhesion to the biomaterial surface. All experiments were conducted with the same root mean square angular rotation velocity (29.63 rad/s) and waveform period. The disk was rotated in platelet‐rich bovine plasma for 2 h, with adhesion quantified by confocal microscopy measurements of immunofluorescently labeled bovine platelets. Platelet adhesion under pulsating flow was found to decay exponentially with increasing shear rate. Adhesion levels were found to depend upon peak platelet flux and shear rate, regardless of rotational waveform. In combination with flow measurements, these results may be useful for predicting regions susceptible to thrombus formation within ventricular assist devices.