On-chip high-pressure picoliter injector for pressure-driven flow through porous media.
On-chip high-pressure picoliter injector for pressure-driven flow through porous media.
复制标题
片上高压皮升注射器用于压力驱动流过多孔介质。
作者:
D. Reichmuth;T. Shepodd;B. Kirby
A high-pressure (> 3 MPa) on-chip injector has been developed for microchip applications including HPLC. The mechanical injector is implemented using in situ photopolymerization of fluorinated acrylates inside wet-etched silica microchips. The injector allows reproducible injections as small as 180 pL with < 250 ms duration. The injector operated robustly over 60 days and over 1000 injections. The injector is unique among polymer-based valves as it functions in aqueous, acetonitrile, and mixed buffers at high pressures without detectable leakage.