Dynamic inversion time for improved 3D late gadolinium enhancement imaging in patients with atrial fibrillation.
Dynamic inversion time for improved 3D late gadolinium enhancement imaging in patients with atrial fibrillation.
复制标题
DOI:
10.1002/mrm.25190
复制
发表时间:
2015-02
影响因子:
3.3
通讯作者:
Firmin DN
中科院分区:
文献类型:
--
作者:
Keegan J;Gatehouse PD;Haldar S;Wage R;Babu-Narayan SV;Firmin DN
High resolution 3D late gadolinium enhancement (LGE) imaging is performed with single R-wave gating to minimise lengthy acquisition durations. In patients with atrial fibrillation (AF), heart rate variability results in variable magnetisation recovery between sequence repeats and image quality is often poor. In this study, we implemented and tested a dynamic inversion time scheme designed to reduce sequence sensitivity to heart rate variations. An inversion-prepared 3D segmented gradient echo sequence was modified so that the inversion time (TI) varied automatically from beat-to-beat (dynamic-TI) based on the time since the last sequence repeat. 3D LGE acquisitions were performed in 17 patients prior to RF ablation of persistent AF both with and without dynamic-TI. Qualitative image quality scores, blood signal-to-ghosting ratios (SGRs) and blood-myocardium contrast-to-ghosting ratios (CGRs) were compared. Image quality scores were higher with dynamic-TI than without (2.2 +/− 0.9 versus 1.8 +/− 1.1, p = 0.008) as were blood-myocardium CGRs (13.8 +/− 7.6 versus 8.3 +/− 6.1, p = 0.003) and blood SGRs (19.6 +/− 8.5 versus 13.1 +/− 8.0, p = 0.003). The dynamic-TI algorithm improves image quality of 3D LGE imaging in this difficult patient population by reducing the sequence sensitivity to RR interval variations.