Evaluation of direct axial sample introduction for ion mobility detection after capillary gas chromatography.
Evaluation of direct axial sample introduction for ion mobility detection after capillary gas chromatography.
复制标题
毛细管气相色谱后离子淌度检测直接轴向进样的评估。
DOI:
10.1016/s0021-9673(01)83338-0
复制
发表时间:
1989
期刊:
影响因子:
--
通讯作者:
HillJr,HH
中科院分区:
文献类型:
--
作者:
StLouis,RH;Siems,WF;HillJr,HH
An ion mobility detector has been designed and constructed for direct axial interfacing with capillary gas chromatography. The principle advantages of this detector were the following: (1) Direct concentric introduction of the capillary column into the ionization region, eliminating peak broadening in the transfer line and improving the efficiency with which neutral molecules were swept from the detector. (2) A variable capillary insertion distance, providing a sensitivity/resolution interplay that could be modified in response to the needs of the assay. (3) An inert gas flow external to the drift cylinder, preventing atmospheric impurities from infiltrating the ultratrace detector. Qualitative and quantitative capabilities of the detector were evaluated using standard preparations ofn-hexyl ether.