Wound healing and catheter thrombosis after implantable venous access device placement in 266 breast cancers treated with bevacizumab therapy
Wound healing and catheter thrombosis after implantable venous access device placement in 266 breast cancers treated with bevacizumab therapy
复制标题
DOI:
10.1097/cad.0b013e328349c7bb
复制
发表时间:
2011-11-01
影响因子:
2.3
通讯作者:
Salmon, Remy J.
中科院分区:
文献类型:
--
作者:
Kriegel, Irene;Cottu, Paul H.;Salmon, Remy J.
The aim of this study was to determine, in a population with metastatic breast cancer treated with bevacizumab therapy, the incidence of wound dehiscence after placement of an implantable venous access device (VAD) and to study the risk of catheter thrombosis. This study enrolled all VADs placed by 14 anesthetists between 1 January 2007 and 31 December 2009: 273 VADs in patients treated with bevacizumab therapy and 4196 VADs in patients not treated with bevacizumab therapy. In the bevacizumab therapy group, 13 cases of wound dehiscence occurred in 12 patients requiring removal of the VAD (4.76%). All cases of dehiscence occurred when bevacizumab therapy was initiated less than 7 days after VAD placement. Bevacizumab therapy was initiated less than 7 days after VAD placement in 150 cases (13 of 150: 8.6%). The risk of dehiscence was the same from 0 to 7 days. In parallel, the VAD wound dehiscence rate in patients not receiving bevacizumab therapy was eight of 4197 cases (0.19%) (Fisher's test significant, P