Quantitative Assessment of Variation in CT Parameters on Texture Features: Pilot Study Using a Nonanatomic Phantom

Quantitative Assessment of Variation in CT Parameters on Texture Features: Pilot Study Using a Nonanatomic Phantom
复制标题

DOI:
10.3174/ajnr.a5139
复制
发表时间:
2017-05-01
影响因子:
3.5
通讯作者:
Sakai, O.
Sakai, O.
中科院分区:
医学2区
文献类型:
--
作者:
Buch, K.;Li, B.;Sakai, O.

文献摘要

被引文献

相似文献

我们的目的是评估的纹理特征的变化的基础上的CT参数的变化的幻影。使用不同的毫安、千伏、切片厚度、间距和采集模式进行扫描。42纹理特征提取使用内部开发的Matlab程序。进行双尾t检验和假检测分析,基于探测器阵列配置(Q值= 0.001-0.006)、切片厚度(Q值= 0.0002-0.001)和采集模式(Q值= 0.003-0.006),纹理特征存在显著差异。毫安和千伏的变化没有明显的影响。
Our aim was to evaluate 'changes in texture features based on variations in CT parameters on a phantom. Scans were performed with varying milliampere, kilovolt, section thickness, pitch, and acquisition mode. Forty-two texture features were extracted by using an in-house-developed Matlab program. Two-tailed t tests and false-detection analyses were performed with significant differences in texture features based on detector array configurations (Q values = 0.001-0.006), section thickness (Q values = 0.0002-0.001), and acquisition mode (Q values = 0.003-0.006). Variations in milliampere and kilovolt had no significant effect.