Inorganic selenium speciation using HPLC-ICP-hexapole collision/reaction cell-MS
Inorganic selenium speciation using HPLC-ICP-hexapole collision/reaction cell-MS
复制标题
使用 HPLC-ICP-六极杆碰撞/反应池-MS 进行无机硒形态分析
DOI:
10.1039/b108267c
复制
发表时间:
2002
影响因子:
3.4
通讯作者:
J. Mermet
中科院分区:
文献类型:
--
作者:
S. Mazáñ;N. Gilon;G. Crétier;J. Rocca;J. Mermet
A chromatographic separation procedure based on the use of a porous graphitic carbon stationary phase and formic acid as eluent was employed to separate Se(IV) and Se(VI), inorganic selenium anions. Detection was performed by using an ICP-MS instrument equipped with a screened torch and a collision/reaction cell. The use of helium, dinitrogen oxide and hydrogen as collision gases was studied to improve selenium detection. The different configurations of the detection system were compared in terms of detection limits. Hydrogen gave the best results with detection limits of 12 and 24 ng L−1 for Se(IV) and Se(VI), respectively.