A detailed test of a BSO calorimeter with 100–800 MeV positrons
A detailed test of a BSO calorimeter with 100–800 MeV positrons
复制标题
DOI:
10.1016/j.nima.2012.08.085
复制
发表时间:
2012-12
影响因子:
1.4
通讯作者:
T. Ishikawa;H. Fujimura;R. Hashimoto;S. Kaida;J. Kasagi;R. Kitazawa;S. Kuwasaki;A. Nakamura;K. Nawa;Y. Okada;M. Sato;H. Shimizu;Koutaku Suzuki;Y. Tajima;Shinichiro Takahashi;Y. Tsuchikawa;H. Yamazaki;H. Yoshida
中科院分区:
文献类型:
--
作者:
T. Ishikawa;H. Fujimura;R. Hashimoto;S. Kaida;J. Kasagi;R. Kitazawa;S. Kuwasaki;A. Nakamura;K. Nawa;Y. Okada;M. Sato;H. Shimizu;Koutaku Suzuki;Y. Tajima;Shinichiro Takahashi;Y. Tsuchikawa;H. Yamazaki;H. Yoshida
A performance test has been made for an electro-magnetic calorimeter prototype comprised four identical BSO crystals arranged in a 2×2 matrix by utilizing a positron beam in the energy range from 100 to 800MeV. The size of each crystal is 40×40×210mm3. This is the world's largest BSO single crystal ever used as a photon detector. The obtained energy resolution is (σE/E)2=((1.71%±0.03%)/E)2+(1.12%±0.08%)2at room temperature, where E is the incident positron energy given in GeV. A BGO calorimeter having the same geometry has been employed in the performance test for comparison. The basic scintillation characteristics of BSO crystals measured without the beam are also presented.