Accurate measurement of sample conductivity in a diamond anvil cell with axis symmetrical electrodes and finite difference calculation
Accurate measurement of sample conductivity in a diamond anvil cell with axis symmetrical electrodes and finite difference calculation
复制标题
DOI:
10.1063/1.3624474
复制
发表时间:
2011-08
期刊:
影响因子:
1.6
通讯作者:
Junliang Yang;Gang Peng;Yonghao Han;C. Gao
中科院分区:
文献类型:
--
作者:
Junliang Yang;Gang Peng;Yonghao Han;C. Gao
We report a relatively precise method of conductivity measurement in a diamond anvil cell with axis symmetrical electrodes and finite difference calculation. The axis symmetrical electrodes are composed of two parts: one is a round thin-film electrode deposited on diamond facet and the other is the inside wall of metal gasket. Due to the asymmetrical configuration of the two electrodes, finite difference method can be applied to calculate the conductivity of sample, which can reduce the measurement error.