A second generation apparatus for time-resolved electron cryo-microscopy using stepper motors and electrospray
A second generation apparatus for time-resolved electron cryo-microscopy using stepper motors and electrospray
复制标题
DOI:
10.1016/j.jsb.2003.09.027
复制
发表时间:
2003-10-01
影响因子:
3
通讯作者:
Trinick, J
中科院分区:
文献类型:
--
作者:
White, HD;Thirumurugan, K;Trinick, J
We describe here a second generation apparatus for studying transient reaction conformations in macromolecules and their complexes by electron cryo-microscopy. Reactions are trapped by rapid freezing in times ranging from a few milliseconds to tens of seconds after initiation. Blotting of the electron microscope grid and freezing it in liquid ethane uses computer controlled micro-stepping motors. For the fastest time resolution, a blotted grid containing a thin film of one reactant is sprayed with small droplets containing a second reactant just before freezing. The spray is produced electrically (electrospray), which gives a dense cloud of droplets < 1 mum in diameter from the 1-2 mul of solution required per grid. A second method in which two solutions are first mixed by turbulent flow and then blotted prior to freezing is used for reactions with time courses > 1 s. (C) 2003 Elsevier Inc. All rights reserved.