Exploring frontiers of orbitrap performance for long transients
Exploring frontiers of orbitrap performance for long transients
复制标题
DOI:
10.1016/j.ijms.2021.116607
复制
发表时间:
2021-05-04
影响因子:
1.8
通讯作者:
Makarov, Alexander
中科院分区:
文献类型:
--
作者:
Denisov, Eduard;Damoc, Eugen;Makarov, Alexander
Experiments on a modified Orbitrap Exploris instrument have shown that even in a compact instrument it is possible to achieve a resolving power in excess of 2,000,000 when appropriate tolerance and tuning requirements are met. Such levels are achieved for a 4-s detection time.A new Orbitrap design with improved pumping allowed, for the first time, isotopic resolution of intact monoclonal antibodies even when using standard nitrogen supply in the Ion Routing Multipole. Longer transients further improve signal-to-noise ratios, approaching levels needed for single-charge resolution in charge detection mass spectrometry. Further steps for improvement of instrumentation are discussed.(c) 2021 Elsevier B.V. All rights reserved.