Direct detection effect in small volume hot electron bolometer mixers
Direct detection effect in small volume hot electron bolometer mixers
复制标题
DOI:
10.1063/1.1887812
复制
发表时间:
2005-04
影响因子:
4
通讯作者:
J. Baselmans;A. Baryshev;S. Reker;M. Hajenius;Jiansong Gao;T. Klapwijk;Y. Vachtomin;S. Maslennikov;S. Antipov;B. Voronov;G. Gol'tsman
中科院分区:
文献类型:
--
作者:
J. Baselmans;A. Baryshev;S. Reker;M. Hajenius;Jiansong Gao;T. Klapwijk;Y. Vachtomin;S. Maslennikov;S. Antipov;B. Voronov;G. Gol'tsman
We measure the direct detection effect in a small volume (0.15μm×1μm×3.5nm) quasioptical NbN phonon cooled hot electron bolometer mixer at 1.6THz. We find that the small signal sensitivity of the receiver is underestimated by 35% due to the direct detection effect and that the optimal operating point is shifted to higher bias voltages when using calibration loads of 300K and 77K. Using a 200GHz bandpass filter at 4.2K the direct detection effect virtually disappears. This has important implications for the calibration procedure of these receivers in real telescope systems.