Interface pressures produced by two different types of lymphedema therapy devices

Interface pressures produced by two different types of lymphedema therapy devices
复制标题

DOI:
10.2522/ptj.20060386
复制
发表时间:
2007-10-01
期刊:
影响因子:
3.2
通讯作者:
Mayrovitz, Harvey N.
Mayrovitz, Harvey N.
中科院分区:
医学2区
文献类型:
--
作者:
Mayrovitz, Harvey N.

文献摘要

被引文献

相似文献

Background and PurposeSequential compression is used to manage lymphedema, but little is known about pressures delivered to the therapeutic targets. This study characterized actual pressures delivered by a traditional compression pump (Lympha Press [LP]) and one using an alternate compression pattern (Flexitouch [FT]).SubjectsTen adults who were healthy volunteered to participate in the study.MethodsPressure-time along the forearm was measured using a 256-pressure sensor array during the pressure cycling of each device. Device assessments were separated by at least 48 hours.ResultsPressure patterns and magnitudes produced by the 2 devices differed considerably. The FT pressure pattern displayed a rapid rise and fall, progressing from the wrist toward the elbow. The LP pressure rose slower and was sustained at a higher level during its inflation cycle. Pressures delivered with the LP were significantly greater than those delivered with the FT.Discussion and ConclusionThe pressure patterns and magnitudes on treated limbs depend on the device. These differences should be considered before selecting a device for a specific patient.