Abdominal-Lead Fetal Electrocardiographic R-Wave Enhancement for Heart Rate Determination

Abdominal-Lead Fetal Electrocardiographic R-Wave Enhancement for Heart Rate Determination
复制标题

DOI:
10.1109/tbme.1980.326631
复制
发表时间:
1980-05
影响因子:
4.6
通讯作者:
S. Azevedo;R. L. Longini
S. Azevedo;R. L. Longini
中科院分区:
工程技术2区
文献类型:
--
作者:
S. Azevedo;R. L. Longini

文献摘要

被引文献

相似文献

胎儿心率(FHR)从外部腹导联心电图波形的可靠性取决于监测设备的R波检测能力。假设ECG的母体部分被其他地方描述的技术取消[1]-[4],我们剩下的胎儿心电图(FECG)可能会被高噪声电平干扰。本文介绍了几种数字信号处理技术,通过这些技术改进了胎儿R波的检测。这些技术-窄带通滤波,局部峰值差分,自回归滤波和匹配滤波-理论上的发展。每种技术的性能测量和分析两个样本胎儿ECG输入。
The reliability of fetal heart rate (FHR) from external abdominal-lead electrocardiographic waveforms depends upon the R-wave detection capability of the monitoring device. Assuming that the maternal component of the ECG is canceiled by techniques described elsewhere [1]-[4], we are left with a fetal electrocardiogm (FECG) which may be conrupted by high noise levels. This paper describes several digital signal processing techniques by which the detection of fetal R-waves is improved. These techniques-narrow band-pass filtering, local peak differencing, autoregressive filtering, and matched filtering-are theoretically developed. The performance of each technique is measured and analyzed for two sample fetal ECG inputs.