Treatment planning of intracranial targets on MRI derived substitute CT data

Treatment planning of intracranial targets on MRI derived substitute CT data
复制标题

DOI:
10.1016/j.radonc.2013.04.028
复制
发表时间:
2013-07-01
影响因子:
5.7
通讯作者:
Nyholm, Tufve
Nyholm, Tufve
中科院分区:
医学1区
文献类型:
--
作者:
Jonsson, Joakim H.;Johansson, Adam;Nyholm, Tufve

文献摘要

被引文献

相似文献

背景资料:磁共振成像(MRI)作为计算机断层扫描(CT)的补充,在放射治疗的目标定义程序的使用正在增加。为了消除由于图像配准引起的系统不确定性,完全基于MRI的工作流程可能是优选的。在目前的试点研究中,我们调查的剂量计算精度自动生成的替代CT(S-CT)图像的头部MRI的基础上。我们还产生数字重建X线片(DRR)从S-CT数据,以评估患者定位的可行性MR images.Methods和材料的基础上:5例患者被纳入本研究。对未经不均匀性校正的CT、s-CT、s-CT数据和体密度指定的MRI图像进行剂量计算。使用点剂量和剂量体积直方图(DVH)的比较,和伽马指数evaluation.Results:结果表明,S-CT图像提高剂量计算的准确性相比,非均匀性校正的剂量计算的方法(平均改善2.0%点),它执行几乎相同的堆密度分配的方法。的S-CT为基础的DRR似乎是足够的病人定位颅内目标,虽然需要进一步调查,在这个subject.Conclusion:的S-CT方法是非常快的,并产生的数据,可用于治疗计划,而不会牺牲准确性。(C)2013爱思唯尔爱尔兰有限公司版权所有。
Background: The use of magnetic resonance imaging (MRI) as a complement to computed tomography (CT) in the target definition procedure for radiotherapy is increasing. To eliminate systematic uncertainties due to image registration, a workflow based entirely on MRI may be preferable. In the present pilot study, we investigate dose calculation accuracy for automatically generated substitute CT (s-CT) images of the head based on MRI. We also produce digitally reconstructed radiographs (DRRs) from s-CT data to evaluate the feasibility of patient positioning based on MR images.Methods and materials: Five patients were included in the study. The dose calculation was performed on CT, s-CT, s-CT data without inhomogeneity correction and bulk density assigned MRI images. Evaluation of the results was performed using point dose and dose volume histogram (DVH) comparisons, and gamma index evaluation.Results: The results demonstrate that the s-CT images improve the dose calculation accuracy compared to the method of non-inhomogeneity corrected dose calculations (mean improvement 2.0% points) and that it performs almost identically to the method of bulk density assignment. The s-CT based DRRs appear to be adequate for patient positioning of intra-cranial targets, although further investigation is needed on this subject.Conclusion: The s-CT method is very fast and yields data that can be used for treatment planning without sacrificing accuracy. (C) 2013 Elsevier Ireland Ltd. All rights reserved.